A Bootstrap Method for Gain Calibration and Resolution Determination of a Lead-Glass Calorimeter
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment(2006)
Department of Physics
Abstract
We describe a method for calibration of a lead-glass calorimeter that does not require a beam of known energy. The technique was used to calibrate the radphi lead-glass calorimeter at Jefferson Lab. The technique described can be applied to any segmented electromagnetic calorimeter capable of detecting all-photon decays of mesons, for example, π0→2γ, η→2γ or ω→π0γ. We also demonstrate how the measured 2γ mass width of the π0 and η mesons can be unfolded to extract the single-shower energy and position resolution functions of the calorimeter.
MoreTranslated text
Key words
meson,photoproduction,radiative,decay,Phi,Radphi,calorimeter,calibration
PDF
View via Publisher
AI Read Science
Must-Reading Tree
Example

Generate MRT to find the research sequence of this paper
Data Disclaimer
The page data are from open Internet sources, cooperative publishers and automatic analysis results through AI technology. We do not make any commitments and guarantees for the validity, accuracy, correctness, reliability, completeness and timeliness of the page data. If you have any questions, please contact us by email: report@aminer.cn