Multiple-Instrument Evaluation of the Consistency and Long-Term Stability of Tip Width Calibration for Critical Dimension Atomic Force Microscopy

JOURNAL OF MICRO-NANOLITHOGRAPHY MEMS AND MOEMS(2017)

NIST

Cited 1|Views3
Abstract
Abstract. Since 2004, standards for calibration of critical dimension atomic force microscope (CD-AFM) tip width have been available both commercially and through the National Metrology Institutes, such as the National Institute of Standards and Technology in the United States. There have been interlaboratory and intermethod comparisons performed on such samples, but less attention has been paid to the long-term stability of standards and monitoring for damage, wear, or contamination. Using three different CD-AFM instruments, we have tested the consistency and long-term stability of two independent reference calibrations for CD-AFM tip width. Both of these tip width calibrations were based on independently implemented transmission electron microscope reference measurements. There were circumstances in which damage occurred or samples needed to be cleaned. Nevertheless, our results show agreement within the uncertainties and stability over a period exceeding 10 years.
More
Translated text
Key words
critical dimension atomic force microscope,linewidth metrology,standards,traceability
PDF
Bibtex
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Data Disclaimer
The page data are from open Internet sources, cooperative publishers and automatic analysis results through AI technology. We do not make any commitments and guarantees for the validity, accuracy, correctness, reliability, completeness and timeliness of the page data. If you have any questions, please contact us by email: report@aminer.cn
Chat Paper
Summary is being generated by the instructions you defined